DQ |
QC results not applicable because of sample matrix |
DR |
Spiked amount less than five times background level |
DS |
Batch Quality Assurance data from another project |
DT |
Method detection limit elevated |
DU |
Insufficient sample quantity for matrix spike/dup matrix spike |
DV |
Corr. coef. for Method of Standard Addition is less than 0.995 |
DW |
Sample result is less than reported value |
DX |
Value < lowest standard (MQL), but > than MDL |
DY |
Compound identified is an analysis at a secondary dilution factor |
DZ |
The positive result is an atypical pattern for diesel analysis |
E |
EPA Flag - Analyte exceeded the concentration range of the GC/MS |
EA |
The positive result is an atypical pattern for gasoline analysis |
EB |
Value is estimated |
EC |
Value determined by the Method of Standard Addition |
ED |
Concentration of total analyte ND; therefore this analyte is ND |
EF |
Compound quantitated at a different dilution |
EG |
Compound quantitated at a 2x dilution factor |
EH |
Compound quantitated at a 5x dilution factor |
EI |
Compound quantitated at a 10x dilution factor |
EJ |
Compound quantitated at a 20x dilution factor |
EK |
Compound quantitated at a 50x dilution factor |
EL |
Compound quantitated at a 100x dilution factor |
EM |
Compound quantitated at a 200x dilution factor |
EN |
Compound quantitated at a 500x dilution factor |
EO |
Compound quantitated at a 1000x dilution factor |
EP |
Compound quantitated at > 1000x dilution factor |
EQ |
Holding time is 28 days by anal. NO3-NO2 on pres. split & sub NO2 |
ER |
Thiocyanate not analyzed separately, free CN+SCN below RL free C |
ES |
Value undetected at the MDL |
ET |
Sample was extracted past end of recommended max. holding time |
EU |
LCS is outside of acceptance limits. MS/MDS are acceptable, no correction |
EV |
Modified method; see narrative |
EW |
Surr. recovery outside of acceptance limits, spike recov. accept. |
EX |
Matrix spike diluted to not detectable during analysis |
EY |
Result exceeds normal dynamic range; reported as a min. est. |
EZ |
Peak coelution, sample concentration may be artificially high |
F1 |
MS and/or MSD recovery exceeds control limits |
FA |
Peak coelution, sample concentration may be artificially low |
FB |
QC ion deviation > 20% relative to labeled std. used for quant. |
FC |
Quantitative interference |
FD |
Interferences affect 10% or more of the total PCDD/PCDF peak area |
FE |
Carbon-labeled int. std. with s/n ratio of < 10:1 |
FF |
S/N ratio 10:1 for the int. std. and assoc. analytes |
FG |
Results rejected due to int. std. rec. outside QC limits |
FH |
Results are valid even though int. std. rec. outside QC limits |
FI |
Analyte in field sample and associated blank |
FJ |
Total DPE contribution to the total PCDF value > 10% |
FK |
EMPC values could be overestimated due to coelution |
FL |
Int. std. ion-abundance ratio measured out of accept. range |
FM |
Poorly resolved GC peak(s) |
FN |
Large closely eluting peak cannot be separated from result |
FO |
Est. maximum possible concentration (EMPC) |
FP |
Extractable hydrocarbons |
FQ |
Gasoline range organics calculated as JP-4 jet fuel |
FR |
Diesel range organics, calculated as JP-4 jet fuel |
FS |
Diesel range organics, as diesel |
FT |
Methanol extraction |
FU |
Heated purge |
FV |
Purge volume 5 mL |
FW |
Purge volume 25 mL |
FX |
Analyte present in the instrument blank |
FY |
Chromatogram not typical of calibration standard |
FZ |
Identified by MS only; based on 1:1 response w/ internal stnd. |
GA |
Components not separable by this method; quantified together |
GB |
Matrix spike recovery not within control limits |
GC |
Second column confirmation performed |
GD |
Secondary column result |
GE |
Post-digestion spike out of control limits |
GF |
Sample absorbance is less than 50% of spike absorbance |
GG |
Analytical results not reliable for this common lab contaminant |
GH |
Reporting limit must be considered an approximation |
GI |
Analyte confirmed by GC/MS |
GJ |
Result determined using SIM mode |
GK |
Quantitated as an isomer pair |
GL |
2,3,7,8-TCDF results have been confirmed on a DB-225 column |
GM |
Result calc. by avg. response factor from init. calib. curve |
GN |
Surrogate recovery is outside of control limits |
GO |
Result est. below lower calib. limit but above target limit |
GP |
Duplicate analysis not within control limits |
GQ |
Post-digest. spike recov. 40% - 85% due to matrix interference |
GR |
Internal standard recovery is outside method recovery limit |
GS |
Reporting limit(s) raised: high level target analyte in sample |
GT |
Post-digestion spike recovery <40% due to matrix interference |
GU |
>10% breakthrough in second carbon col. due to matrix effects |
GV |
Result detected is below the lowest standard and above zero |
GW |
Post-digestion spike recovery between 115%-150% |
GX |
All reporting limits raised due to high conc. dissolved solids |
GY |
Analyte assoc. with sample processing & analysis in lab environ. |
GZ |
Secondary ion used for quantitation |
H |
A holding time violation has occurred. |
HA |
Quantit. of unknown hydrocarbon(s) in sample based on diesel |
HB |
Quantit. of unknown hydrocarbon(s) in sample based on gasoline |
HC |
Analyte is this compound or an isomer |
HD |
Chromat. profile inconsistent with pattern(s) of ref. fuel stnds. |
HE |
Extract was run through separation method; see case narrative |
HF |
SW3630B clean-up method performed |
HG |
Result exceeds calib. range, but not lab's; considered valid |
HH |
Result exceeds linear range; concentration may be understated |
HI |
SW3611B clean-up method performed |
HJ |
Blank spike recovery not within control limits |